Fluke 718 100G Pressure Calibrator

The Fluke 718 pressure calibrator family (includes 4 models) provide a total pressure calibration solution for transmitters, gauges and switches. This compact pressure calibration solution is about a third of the size of comparable instruments and weighs just one kilogram (two pounds). The unique easy to clean pump helps protect the pump from damage and provides the ability to service the pump without disassembly, even in the field.



Description

Technical Specifications
Standard
718 Series
Measure Pressure  (internal sensor)
Accuracy  0.025% full scale
Functions  zero, min/max, hold, damp
Note:  Dry air, non-corrosives only
718 1G
Range:  -1 PSI to +1 PSI, -68.9 mbar to 68.9 mbar (-6.89 kPa to 6.89 kPa)
Resolution:  0.0001 psi, 0.001 mbar, 0.0001 kPa
Over Pressure:  5x Full Scale
Functions:  Zero, Min, Max, Hold, Damp
718 30G
Range:  -12 PSI to 30 PSI, (-850 mbar to 2 bar, -85 to 206.84 kPa)
Resolution:  0.001 psi, 0.1 mbar, 0.01 kPa
Over Pressure:  5x Full Scale
Functions:  Zero, Min, Max, Hold, Damp
718 100G
Range:  -12 PSI to 100 PSI, (-850 mbar to 7 bar, -85 to 689.48 kPa)
Resolution:  0.01 psi, 0.1 mbar, 0.01 kPa
Over Pressure:  2x Full Scale
Functions:  Zero, Min, Max, Hold, Damp
718 300G
Range:  -12 PSI to 300 PSI, (-850 mbar to 20 bar, -85 to 2068.42 kPa )
Resolution:  0.01 psi, 1 mbar, 0.1 kPa
Over Pressure:  375 PSI 25 bar
Functions:  Zero, Min, Max, Hold, Damp
Fluke 718 series via using 700 Series Pressure Modules
Range:  29 pressure modules, 0 – 1 in H20 to 10,000 psi, 2.5 mbar to 700 bar. Over pressure per pressure module specs
Resolution:  Per pressure module specs
Accuracy:  To 0.025 % of full span, per pressure module specs
Functions:  Zero, Min, Max, Hold, Damp
 Note: media compatibility per pressure module specs
Fluke 718 series via built in pump
Range:  -12 PSI or -850 mbar to full scale Supported Pressure Units; psi, in H20 (4 °C), in H20 (20 °C), cm H20 (4 °C), cm H20 (20 °C), bar, mbar, kPa, inHg, mmHg, kg/cm
Loop Power:  Range: 24 V dc
 Accuracy: +/- 10 %
 Note: Drive; 20 mA into 1000 Ω for battery > 6.8V; 700 Ω for battery 5.8 to 6.8 V
mA Measure:  Range: 0 mA to 24 mA
 Resolution: 0.001 mA
 Accuracy: 0.015 % + 1 count

 

General Specifications
Maximum voltage
 30 volts: Non-operating
Temperature
 -40 °C to 60 °C
Operating Temperature
 -10 °C to 55° C
Relative Humidity
 95 % (10 °C to 30 °C); 75 % (30 °C to 40 °C); 45 % (40 °C to 50 °C); 35 % (50 °C to 55 °C)
Operating Altitude
 3,000 M max
Shock
 1 M drop test
Vibration
 Random, 2 g, 5 Hz – 500 Hz
Safety
 CSA C22.2 No. 1010.1: 1992
EMC
 EN50082-1: 1992 and EN55022: 1994 Class B
Size
 216 mm x 94 mm x 66 mm (8.50 in x 3.72 in x 2.60 in)
Weight
 992 g (35 oz)
Power, battery life
 9V battery ANSI/NEDA 1604A or IEC 6LR619V alkaline; two batteries in 718
Battery Life
 4 to 20 hours typical, depending on functions used
Warranty
 Three years (one year on pressure pump in Fluke 718)
Display
 LCD, 5 digit pressure and current simultaneous
Note
 For use with non-corrosive gasses
Fluke 718 100G Pressure Calibrator